Optical coherence tomography (OCT)
Optical coherence tomography (OCT) is an optical signal acquisition and processing method. It captures micrometer-resolution, three-dimensional images from within optical scattering media (e.g., biological tissue). Optical coherence tomography is an interferometric technique, typically employing near-infrared light. The use of relatively long wavelength light allows it to penetrate into the scattering medium. Confocal microscopy, another similar technique, typically penetrates less deeply into the sample.
Depending on the properties of the light source (superluminescent diodes, ultrashort pulsed lasers and supercontinuum lasers have been employed), optical coherence tomography has achieved sub-micrometer resolution (with very wide-spectrum sources emitting over a ~100 nm wavelength range).
Optical coherence tomography is one of a class of optical tomographic techniques. A relatively recent implementation of optical coherence tomography, frequency-domain optical coherence tomography, provides advantages in signal-to-noise ratio, permitting faster signal acquisition. Commercially available optical coherence tomography systems are employed in diverse applications, including art conservation and diagnostic medicine, notably in ophthalmology where it can be used to obtain detailed images from within the retina. Recently it has also begun to be used in interventional cardiology to help diagnose coronary artery disease.
Introduction:
Starting
from white-light interferometry for in vivo ocular eye
measurements imaging of biological tissue,
especially of the human eye, was investigated by multiple groups worldwide. A
first two-dimensional in vivo depiction of a human eye fundus
along a horizontal meridian based on white light interferometric depth scans
was presented at the ICO-15 SAT conference in 1990. Further developed in 1990 by
Naohiro Tanno, then a professor at Yamagata
University, and in particular since 1991 by Huang et al., optical coherence tomography
(OCT) with micrometer resolution and cross-sectional imaging capabilities has
become a prominent biomedical tissue-imaging technique; it is particularly
suited to ophthalmic applications and other tissue imaging requiring micrometer
resolution and millimeter penetration depth. First in vivo OCT
images – displaying retinal structures – were published in 1993. OCT has also been used for
various art conservation projects, where it is used to analyze different layers
in a painting. OCT has critical advantages over other medical imaging systems. Medical
ultrasonography, magnetic
resonance imaging (MRI) and confocal microscopy are
not suited to morphological tissue imaging: the first two have poor resolution;
the last lacks millimeter penetration depth.
OCT bases
itself upon low coherence
interferometry. In conventional interferometry
with long coherence length (laser interferometry), interference of light occurs
over a distance of meters. In OCT, this interference is shortened to a distance
of micrometers, thanks to the use of broadband light sources (sources that can
emit light over a broad range of frequencies). Light with broad bandwidths can
be generated by using superluminescent
diodes(superbright LEDs) or lasers with extremely short pulses (femtosecond lasers). White light is also a
broadband source with lower power.
Light in
an OCT system is broken into two arms—a sample arm (containing the item of
interest) and a reference arm (usually a mirror). The combination of reflected
light from the sample arm and reference light from the reference arm gives rise
to an interference pattern, but only if light from both arms have travelled the
"same" optical distance ("same" meaning a difference of
less than a coherence length). By scanning the mirror in the reference arm, a
reflectivity profile of the sample can be obtained (this is time domain OCT).
Areas of the sample that reflect back a lot of light will create greater
interference than areas that don't. Any light that is outside the short
coherence length will not interfere. This reflectivity profile, called an A-scan, contains information about the spatial dimensions
and location of structures within the item of interest. A cross-sectional
tomograph (B-scan) may be achieved by laterally combining
a series of these axial depth scans (A-scan). En face imaging (C-scan)
at an acquired depth is possible depending on the imaging engine used.
Optical
Coherence Tomography, or ‘OCT’, is a technique for obtaining sub-surface images
of translucent or opaque materials at a resolution equivalent to a low-power
microscope. It is effectively ‘optical ultrasound’, imaging reflections from
within tissue to provide cross-sectional images.
OCT is
attracting interest among the medical community, because it provides tissue
morphology imagery at much higher resolution (better than 10 µm) than
other imaging modalities such as MRI or ultrasound.
The key
benefits of OCT are:
§ Live sub-surface images at near-microscopic resolution
§ Instant, direct imaging of tissue morphology
§ No preparation of the sample or subject
§ No ionizing radiation
OCT
delivers high resolution because it is based on light, rather than sound or
radio frequency. An optical beam is directed at the tissue, and a small portion
of this light that reflects from sub-surface features is collected. Note that
most light is not reflected but, rather, scatters off at large angles. In
conventional imaging, this diffusely scattered light contributes background
that obscures an image. However, in OCT, optical coherence is used to record
the optical path length of received photons allowing rejection of most photons
that scatter multiple times before detection. Thus OCT can build up clear 3D
images of thick samples by rejecting background signal while collecting light
directly reflected from surfaces of interest.
Within the
range of noninvasive three-dimensional imaging techniques that have been
introduced to the medical research community, OCT as an echo technique is
similar to ultrasound imaging.
Other medical imaging techniques such as computerized axial tomography,
magnetic resonance imaging, or positron emission tomography do not utilize the
echo-location principle.
The
technique is limited to imaging 1 to 2 mm below the surface in biological
tissue, because at greater depths the proportion of light that escapes without
scattering is too small to be detected. No special preparation of a biological
specimen is required, and images can be obtained ‘non-contact’ or through a
transparent window or membrane. It is also important to note that the laser
output from the instruments is low – eye-safe near-infra-red light is used –
and no damage to the sample is therefore likely.
Theory
The principle OCT is white light or low
coherence interferometry. The optical setup typically consists of an
interferometer (Fig. 1, typically Michelson type)
with a low coherence, broad bandwidth light source. Light is split into and
recombined from reference and sample arm, respectively.
Time domain OCT
In time
domain OCT the pathlength of the reference arm is translated longitudinally in
time. A property of low coherence interferometry is that interference, i.e. the
series of dark and bright fringes, is only achieved when the path difference
lies within the coherence length of the light source. This interference is
called auto correlation in a symmetric interferometer (both arms have the same
reflectivity), or cross-correlation in the common case. The envelope of this
modulation changes as pathlength difference is varied, where the peak of the
envelope corresponds to pathlength matching.
The interference of two partially coherent light beams can be expressed in terms of the source intensity, , as
- where represents the interferometer beam splitting ratio, and is called the complex degree of coherence, i.e. the interference envelope and carrier dependent on reference arm scan or time delay , and whose recovery of interest in OCT. Due to the coherence gating effect of OCT the complex degree of coherence is represented as a Gaussian function expressed as[15]where represents the spectral width of the source in the optical frequency domain, and is the centre optical frequency of the source. In equation (2), the Gaussian envelope is amplitude modulated by an optical carrier. The peak of this envelope represents the location of sample under test microstructure, with an amplitude dependent on the reflectivity of the surface. The optical carrier is due to the Doppler effect resulting from scanning one arm of the interferometer, and the frequency of this modulation is controlled by the speed of scanning. Therefore translating one arm of the interferometer has two functions; depth scanning and a Doppler-shifted optical carrier are accomplished by pathlength variation. In OCT, the Doppler-shifted optical carrier has a frequency expressed aswhere is the central optical frequency of the source, is the scanning velocity of the pathlength variation, and is the speed of light.The axial and lateral resolutions of OCT are decoupled from one another; the former being an equivalent to the coherence length of the light source and the latter being a function of the optics. The coherence length of a source and hence the axial resolution of OCT is defined asFrequency domain OCT (FD-OCT)In frequency domain OCT the broadband interference is acquired with spectrally separated detectors (either by encoding the optical frequency in time with a spectrally scanning source or with a dispersive detector, like a grating and a linear detector array). Due to the Fourier relation (Wiener-Khintchine theorem between the auto correlation and the spectral power density) the depth scan can be immediately calculated by a Fourier-transform from the acquired spectra, without movement of the reference arm. This feature improves imaging speed dramatically, while the reduced losses during a single scan improve the signal to noise proportional to the number of detection elements. The parallel detection at multiple wavelength ranges limits the scanning range, while the full spectral bandwidth sets the axial resolution.Spatially encoded frequency domain OCT (spectral domain or Fourier domain OCT)SEFD-OCT extracts spectral information by distributing different optical frequencies onto a detector stripe (line-array CCD or CMOS) via a dispersive element (see Fig. 4). Thereby the information of the full depth scan can be acquired within a single exposure. However, the large signal to noise advantage of FD-OCT is reduced due the lower dynamic range of stripe detectors in respect to single photosensitive diodes, resulting in an SNR (signal to noise ratio) advantage of ~10 dB at much higher speeds. This is not much of a problem when working at 1300 nm, however, since dynamic range is not a serious problem at this wavelength range.The drawbacks of this technology are found in a strong fall-off of the SNR, which is proportional to the distance from the zero delay and a sinc-type reduction of the depth dependent sensitivity because of limited detection linewidth. (One pixel detects a quasi-rectangular portion of an optical frequency range instead of a single frequency, the Fourier-transform leads to the sinc(z) behavior). Additionally the dispersive elements in the spectroscopic detector usually do not distribute the light equally spaced in frequency on the detector, but mostly have an inverse dependence. Therefore the signal has to be resampled before processing, which can not take care of the difference in local (pixelwise) bandwidth, which results in further reduction of the signal quality. However, the fall-off is not a serious problem with the development of new generation CCD or photodiode array with a larger number of pixels.Synthetic array heterodyne detection offers another approach to this problem without the need for high dispersion.Time encoded frequency domain OCT (also swept source OCT)TEFD-OCT tries to combine some of the advantages of standard TD and SEFD-OCT. Here the spectral components are not encoded by spatial separation, but they are encoded in time. The spectrum either filtered or generated in single successive frequency steps and reconstructed before Fourier-transformation. By accommodation of a frequency scanning light source (i.e. frequency scanning laser) the optical setup (see Fig. 5) becomes simpler than SEFD, but the problem of scanning is essentially translated from the TD-OCT reference-arm into the TEFD-OCT light source. Here the advantage lies in the proven high SNR detection technology, while swept laser sources achieve very small instantaneous bandwidths (=linewidth) at very high frequencies (20–200 kHz). Drawbacks are the nonlinearities in the wavelength, especially at high scanning frequencies. The broadening of the linewidth at high frequencies and a high sensitivity to movements of the scanning geometry or the sample (below the range of nanometers within successive frequency steps).
Scanning Schemes :Focusing the light beam to a point on the surface of the sample under test, and recombining the reflected light with the reference will yield an interferogram with sample information corresponding to a single A-scan (Z axis only). Scanning of the sample can be accomplished by either scanning the light on the sample, or by moving the sample under test. A linear scan will yield a two-dimensional data set corresponding to a cross-sectional image (X-Z axes scan), whereas an area scan achieves a three-dimensional data set corresponding to a volumetric image (X-Y-Z axes scan), also called full-field OCT.Single point (confocal) OCTSystems based on single point, or flying-spot time domain OCT, must scan the sample in two lateral dimensions and reconstruct a three-dimensional image using depth information obtained by coherence-gating through an axially scanning reference arm (Fig. 2). Two-dimensional lateral scanning has been electromechanically implemented by moving the sample using a translation stage, and using a novel micro-electro-mechanical system scanner.Parallel (or full field) OCTParallel OCT using a charge-coupled device (CCD) camera has been used in which the sample is full-field illuminated and en face imaged with the CCD, hence eliminating the electromechanical lateral scan. By stepping the reference mirror and recording successive en face images a three-dimensional representation can be reconstructed. Three-dimensional OCT using a CCD camera was demonstrated in a phase-stepped technique, using geometric phase-shifting with a Linnik interferometer, utilising a pair of CCDs and heterodyne detection, and in a Linnik interferometer with an oscillating reference mirror and axial translation stage. Central to the CCD approach is the necessity for either very fast CCDs or carrier generation separate to the stepping reference mirror to track the high frequency OCT carrier.Smart detector array for parallel TD-OCTA two-dimensional smart detector array, fabricated using a 2 µm complementary metal-oxide-semiconductor (CMOS) process, was used to demonstrate full-field OCT. Featuring an uncomplicated optical setup (Fig. 3), each pixel of the 58x58 pixel smart detector array acted as an individual photodiode and included its own hardware demodulation circuitry.
Optical coherence tomography is an established medical imaging technique. It is widely used, for example, to obtain high-resolution images of the retinaand the anterior segment of the eye, which can, for example, provide a straightforward method of assessing axonal integrity in multiple sclerosis.Researchers also seek to develop a method that uses frequency domain OCT to image coronary arteries in order to detect vulnerable lipid-rich plaques. Furthermore, studies by Dr. Thomas Milner and Dr. Grady Rylander at the University of Texas in Austin suggest its use as a valuable tool in early-stage detection of glaucoma.Optical coherence tomography is also applicable and increasingly used in industrial applications, such as Non Destructive Testing(NDT), material thickness measurements, and in particular thin silicon wafers and compound semiconductor wafers thickness measurements surface roughness characterization, surface and cross-section imaging and volume loss measurements. OCT systems with feedback can be used to control manufacturing processes. With high speed data acquisition, and sub-micron resolution, OCT is adaptable to perform both inline and off-line. Fiber-based OCT systems are particularly adaptable to industrial environments.These can access and scan interiors of hard-to-reach spaces, and are able to operate in hostile environments - whether radioactive, cryogenic or very hot.Optical coherence tomography is an established medical imaging technique. It is widely used, for example, to obtain high-resolution images of the retinaand the anterior segment of the eye, which can, for example, provide a straightforward method of assessing axonal integrity in multiple sclerosis.Researchers also seek to develop a method that uses frequency domain OCT to image coronary arteries in order to detect vulnerable lipid-rich plaques. Furthermore, studies by Dr. Thomas Milner and Dr. Grady Rylander at the University of Texas in Austin suggest its use as a valuable tool in early-stage detection of glaucoma.
Optical coherence tomography is also applicable and increasingly used in industrial applications, such as Non Destructive Testing(NDT), material thickness measurements, and in particular thin silicon wafers and compound semiconductor wafers thickness measurements surface roughness characterization, surface and cross-section imaging and volume loss measurements. OCT systems with feedback can be used to control manufacturing processes. With high speed data acquisition, and sub-micron resolution, OCT is adaptable to perform both inline and off-line. Fiber-based OCT systems are particularly adaptable to industrial environments.These can access and scan interiors of hard-to-reach spaces, and are able to operate in hostile environments - whether radioactive, cryogenic or very hot.